Section 5: Service and Maintenance
08/12 41 51300 Rev. 1
5.2.2. Solutions
For all errors not listed or for help, please call your local Fairbanks Service Representative.
SYMPTOM CAUSE SOLUTION
UL1or UL2
Below ZERO
Bind
Check for platform bind.
Check that the “S” hook is attached.
Check that the hook is under the levers end at
the bottom of the scale.
Check for proper platform assembly.
Possible damaged load cell.
Call for service.
OL1orOL2
Over Capacity
Bind
Possible damaged
load cell.
Possible A/D failure.
Check platform to be empty, then press ZERO.
Check for platform bind.
Call for service.
Does not return to
“0.0”
Bind
Check for platform bind.
Check that the “S” hook is attached.
Check that the hook is under the levers end at
the bottom of the scale.
Check for proper platform assembly.
“Error3”
Bind
Possible damaged
load cell.
Possible A/D failure.
Check for platform bind.
Check that the “S” hook is attached.
Check that the hook is under the levers end at
the bottom of the scale.
Check for proper platform assembly.
Call for service.
Blank display
Battery pack
discharged.
No power at outlet
Batteries dead
Faulty instrument
Replace or Recharge the battery pack.
Check AC outlet; check circuit box.
Call for service.
“LoBAtt”
Batteries low
Recharge the battery pack.