Instructions / Assembly

5
Note: If the equipment is used in a manner not specied by the manufacturer, the protection
provided by the equipment may be impaired.
SAFETY
Features
Emergency Stop The safety switch can be found on either side of the chamber and
can be used at any time to stop the operation. Pressing EMERGENCY
STOP disconnects all power.
Open Door(s) BioShift chamber will not operate when the doors are open. The
doors cannot be opened during cycle time unless EMERGENCY
STOP is pressed. Both doors cannot be opened at the same time
unless EMERGENCY STOP is pressed.
Interlock Logic The clean (controller side) door will remain locked until a cycle is
complete. When complete, the clean door will remain unlocked until
the dirty (start button side) door is opened, or a new cycle is started.
If a cycle is interrupted, it will start over at the beginning. The dirty
door will remain locked anytime the clean door is open. Pressing the
EMERGENCY STOP will unlock both doors.
Start Button Illuminates blue when ready to start cycle.
Chamber The sealed chamber is meant to protect users from UV-C radiation.
Window The tempered glass windows are opaque to UV-C.
Controller The controller has a countdown timer which allows the user to
monitor Lamp Life, replacement recommended after 1000 cycles,
3000 hours, or after a failed Dose Verication Test (see Maintenance
section).