Instructions / Assembly
MAINTENANCE
Cleaning
• Do not use industrial degreasers or strong solvents such as brake cleaners.
• Do not override safety features during maintenance / cleaning.
• When cleaning all reective aluminum surfaces, do not use abrasives, they will
reduce the reectivity.
• Outside: BioShift chamber can be cleaned with a damp towel. Strong solvents
or detergents are not recommended. A typical household cleaning solvent is
recommended.
• Inside: Clean inside chamber walls with ethanol (reagent grade) and a soft towel.
Be careful not to spray or apply chemicals directly on the UV-C lamps. This may
result in total lamp destruction.
• Inspect the aluminum panels, if they appear to be degrading and reducing
reectivity, consider replacing.
UV-C Dose Verification
It is important to regularly verify the UV-C dose output of the BioShift chamber to
ensure proper disinfection. Every 50 cycles or once per month, whichever comes rst,
the operator shall verify that the UV-C dose exceeds a minimum of 100 mJ/cm
2
in a
short, programmed test cycle. Before proceeding with the Dose Verication, ensure the
chamber has completed a full cycle within the last 5 minutes so that the lamps are warmed
up. With an empty chamber, place a card (below) face-up at the center of the tray in the
bottom position. On the touchscreen display, select the “100 mJ/cm
2
Dose Test” button
to start a test cycle. Upon completion of the test cycle, the operator will evaluate the color
development of the center circle. If the center color matches the 100 mJ/cm
2
region, the
chamber has passed the test. If the color does not match 100 mJ/cm
, all the lamps in
the chamber must be replaced. Note: Do not use dose verication as a disinfection cycle.
Replacement 100 mJ/cm
2
test cards (929002100327) are available from ONCE
®
at www.once.lighting/uv-cdosimeter. Do not store cards in direct sunlight.
13