User guide
Table 35
LED Name Description
LAN Constantly green when LAN is operable
Constantly red when LAN is not operable
Flashing green when device
identification is enabled
Constantly orange during boot phase of
the instrument
Power Constantly orange when instrument is
OFF (in power standby mode)
Constantly green when instrument is ON
(powered)
Problems with the M8061A
Problems with the M8061A - Concepts
When you have opened the External Instrument(s) - Config window and have
selected M8061A from the external instruments, select Mux with DeEmphasis
under Enable Function column in the list, the M8061A Connection dialog appears.
Cannot-Connect Problems
An error message may appear when you now press Enable.
Check the Power Supply
Ensure that the power cord is connected. Check for the steady green "STATUS"
light that appears on the ESM of the chassis indicating a power ready status of
the chassis which mounts the M8061A module. Check for the status LEDs present