User guide

The oscilloscope should show the following picture:
If you see only a straight line, ensure that the pattern generator is not disabled and
that it is set up to generate a pattern (e.g. PRBS).
Settings for Clock Multiplier Test
Following are the settings to test the output of the clock multiplier:
Preset Instrument State
Set Data Rate to 3GB/s
Set Clock Offset to 0V
Set Clock Out Amplitude to 400mV
Set PG Trigger Out Offset to 0V
Enable Clock Multiplier in the Config window
Set the following for Clock Multiplier:
Input Frequency 3GHz
Termination ON at 0V
Check the Output of the N4916B (Clock Multiplier Part) Using a
Scope ( Clock Multiplier 1)
Following are the steps to test clock multiplier 1 using a scope:
Solving Problems 9
Agilent J-BERT N4903B High-Performance Serial BERT 507