User guide

Once the de-emphasis function is enabled you can see the de-emphasis
parameters in the PG Output screen.
7
Again, from the Config window, corrosponding to N4916B, clear the De-
Emphasis function check box. It will disable the de-emphasis function.
If you could proceed until here, the connection of the user software to the N4916B
is ok.
Settings for De-Emphasis Tests
The following are the settings to test the output of the de-emphasis:
Preset Instrument State
Set the PRBS 2^7-1
Set the Data Rate to 3GB/s
Set the Data Offset to 0V
Set Data Amplitude to 400mV
Enable De-Emphasis in the Config window
Set Pre Cursor to +12dB
Set Post Cursor 1 to: -6dB
Set Post Cursor 2 to: -3dB
Check the Output of the N4916B (De-Emphasis Part) Using a Scope
1
Connect the scope Data out of N4916B
2
Connect the Trigger In of Scope to Trigger Out of PG
3
Terminate N4916B Data /Out with 50Ohm
4
Terminate PG Trigger/Out with 50Ohm
5
Terminate PG Data/Out with 50Ohm
9 Solving Problems
504 Agilent J-BERT N4903B High-Performance Serial BERT