When using lower target BER values, the required measurement time per Input
Timing Setting is getting longer very quickly.
The required measurement time for a target BER of 10
-9
is approximately 338e6
seconds per Input Timing Setting, which is not practicable in a real measurement
setup.
N4876A System Setup Adjustment
N4876A System Setup Adjustment
The N4876A has to be adjusted to ensure error free operation for all data rates.
This adjustment is required once for a dedicated setup of N4903B, N4876A and
the cable kit being used, and has to be repeated whenever one of these are changed
in the setup or the operating temperature is changed by more than 5
o
Celsius,
compared to the previous adjustment.
N O T E
When the N4876A is enabled from the N4903B, the jitter amplitude in time is
decreased by a factor of 2 to make the jitter amplitude in UI at the N4876A output
match the setting in UI in the N4903B GUI.
Performing System Setup Adjustment Setting
Follow the given steps to perform system setup adjustment setting:
1
From the Navigation Menu, select External Instrument(s) and then click
Config sub-menu. The Config window shows a list of all instruments connected
to the N4903B.
2
Select the external instrument from the given list. For example, here you will
select N4876A.
3
Click Adjustment... to continue.
A message will pop up which warns you to save the instrument settings before
continuing the auto adjustment process.