User guide

Connect the N4903B’s Trigger Output with the oscilloscope’s Front panel
Trigger input
Connect the N4916B’s Data Output with channel 1 of the oscilloscope
Terminate the N4916B’s complement output using a 50Ohm termination
resistor
1
N4903B
a
Open the configuration page for external instruments
b
Enable the De-Emphasis function
c
Click the Load Calibration Setting… button
d
Set the PG’s data rate to 14.2 Gb/s
e
Open the external instruments configuration screen again
2
86100A
a
Enable Channel 1, set the vertical scale to 100 mV/div and the offset voltage
to 0 V
b
Select the appropriate trigger input and set the threshold voltage to 0 V.
Select Trigger Mode ‘Normal’ (DC – 2.5 GHz)
c
Set the time base to 28.2 ps
d
Adjust the Delay to show the start of the 2
7
-1 PRBS at the center of the
screen (0000001111111 bit sequence)
e
Set the Waveform Persistence to infinite
f
Enable Eye Mask Mode
For each Input Timing Setting (1 to 6) execute the following steps:
Set the Input Timing on the N4903B
Clear the oscilloscope’s display
Wait for waveforms to accumulate for approximately 6 minutes
Check the waveform on the oscilloscope for erroneous measurements that
indicate a timing violation at the input of the N4916B
When all 6 settings have been measured, select the optimum input timing setting
(keep in mind that one erroneous sample on the screen has to be treated as an
error).
Example screen shots
Input Timing Setup = 1
Solving Problems 9
Agilent J-BERT N4903B High-Performance Serial BERT 485
Calibration Preparation
Calibration Sequence