User guide

Table 33
Input
Timing
Errors Forward
distance
to ‘bad’
setting
Backwar
d
distance
to ‘bad’
setting
1 0 3 1
2 0 2 2
3 0 1 3
4 50 0 0
5 1000 0 0
6 100 0 0
Case-2 (Calibration using an Oscilloscope)
Connect Data Out of N4916B to the oscilloscope.
Terminate Data Out of N4916B.
Connect Trigger-Out of PG with Trigger-In of oscilloscope.
Terminate Trigger Out of PG with 50 Ohm termination resistor.
Select each of the 6 Input Timing Settings and measure the output of the
N4916B with the oscilloscope.
Measurement has to be done checking for erroneous sampled bits, so the
oscilloscope has to be adjusted to measure at a xx0 to 111 transition of the
pattern.
The oscilloscope has to be configured to display every measured sample
infinitely.
Measurement time has to be at least long enough for the required target
BER (e.g. 1^-12 at a confidence level of 95%).
A measurement is considered ‘error free’ when there are no samples taken
‘outside’ the regular signal trace.
Selection of the optimum Input Timing setting is done as shown above for
the BER measurement.
4
Selection of the optimum Input Timing setting is done as shown above for the
BER measurement.
5
Optionally reload the previously stored instrument state
Solving Problems 9
Agilent J-BERT N4903B High-Performance Serial BERT 483