W A R N I N G
The instrument settings will be lost once you start the calibration process. It is
recommended to save the instrument settings before you start the calibration
process.
5
Click Yes to continue if you have already saved the settings.
Calibration Procedure
The following steps explain the calibration prodecure:
1
Store current instrument setting if the current setting is required after the
calibration is done
2
Load the calibration setting. For more details, refer “Loading Calibration
Settings” on page 480.
3
Depending on if an error detector available or not, we have 2 cases:
Case -1 (Calibration using an Error Detector)
– Connect Data Out of N4916B to Data In of Error Detector.
– Terminate Data Out of N4916B and Data In of Error Detector with 50 Ohm
termination resistor.
– Connect Clock Out of Pattern Generator with Clock In of Error Detector.
– Terminate Clock Out of Pattern Generator with 50 Ohm termination resistor
– For each of the 6 timing settings do a BER measurement using the following
accumulation setting
Solving Problems 9