User guide

W A R N I N G
The instrument settings will be lost once you start the calibration process. It is
recommended to save the instrument settings before you start the calibration
process.
5
Click Yes to continue if you have already saved the settings.
Calibration Procedure
The following steps explain the calibration prodecure:
1
Store current instrument setting if the current setting is required after the
calibration is done
2
Load the calibration setting. For more details, refer “Loading Calibration
Settings” on page 480.
3
Depending on if an error detector available or not, we have 2 cases:
Case -1 (Calibration using an Error Detector)
Connect Data Out of N4916B to Data In of Error Detector.
Terminate Data Out of N4916B and Data In of Error Detector with 50 Ohm
termination resistor.
Connect Clock Out of Pattern Generator with Clock In of Error Detector.
Terminate Clock Out of Pattern Generator with 50 Ohm termination resistor
For each of the 6 timing settings do a BER measurement using the following
accumulation setting
Solving Problems 9
Agilent J-BERT N4903B High-Performance Serial BERT 481