User guide
This opens the Jitter Setup window.
2
Enable jitter generation. Press the respective checkbox.
3
Activate Sinusoidal jitter.
4
Activate Periodic jitter
5
Set the jitter amplitude to 0 UI for both Sinusoidal jitter and Periodic jitter.
N O T E
When rSSC is enabled the measurement does not use SJ. The maximum available
jitter amplitude is reduced for the jitter tolerance measurement.
Running the Test
Once you have specified the Standard, Frequency, and BER, press the Start button
to run the test.
The bar in the lower right-hand corner of the screen shows the progress.
You can also abort the test at any time by pressing the Abort button.
When the measurement is either completed or aborted, you can obtain a list of
measured points by clicking the following tab which appears once the
measurement is stopped.
Clicking on the above tab you get the list of measured points.
Jitter Tolerance Tests 8
Agilent J-BERT N4903B High-Performance Serial BERT 459