User guide

To specify the jitter components:
1
Activate a jitter source (press the corresponding button).
2
Press Edit button to access all parameters of the corresponding jitter type.
If you want to specify the jitter components for Constant Amplitude Sweep,
skip steps 5 to 6.
However, if you want to specify the jitter components for Variable Amplitude
Sweep, skip steps 3 to 4 and go to step 5.
3
Activate Constant Amplitude Sweep (press Constant Amplitude Sweep
button).
This opens a window which allows you to set all parameters available for the
given jitter type.
4
Specify the following parameters:
Amplitude
Frequency Range
Sweep Time
Nr. of Steps
For more information on parameters, see “Periodic Jitter 2 – Constant
Amplitude Sweep Parameters” on page 423.
5
Activate Variable Amplitude Sweep (press Variable Amplitude Sweep button).
Jitter Tolerance Tests 8
Agilent J-BERT N4903B High-Performance Serial BERT 411