one single frame error. In contrast to this, the Serial BERT loses pattern
synchronization and must re-synch before the test can be continued. Currently,
the detection mechanism for loss of synchronization is a high error ratio. This
causes the Serial BERT to show many errors in the "Accumulated Results"
window though there was only one single error. The serial BERT shows with
such DUTs high error number, the number is not accurate. The Auto Resync
Counter events are shown in the Accumulated Results window and the same
are logged in file. Log settings can be changed in Error Detector's Accumulation
Setup dialog.
Accumulation Parameters for 8B/10B Comparison
For detailed information on 8B/10B Comparison accumulation parameters, see
“Accumulation Parameters” on page 382.
Accumulated Measurements for Bit Comparison without PCIe3
SKPOS
The Accumulated Results window for Bit Comparison without PCIe3 SKPOS
contains a graph and two tables:
N O T E
The Accumulated Results window for Bit Comparison without PCIe3 SKPOS does
not contain tables for G.821Measurement, Internal Results and Burst Results.
Evaluating Results 7
Agilent J-BERT N4903B High-Performance Serial BERT 389