User guide

N O T E
During measurement logging, the Serial BERT logs data in ten-second intervals.
Your log file may be missing up to the last ten seconds of data. To avoid this
condition, accumulate for 10 seconds longer than desired.
Accumulated Measurements - Procedures
This section describes how to run accumulated measurements and how to analyze
the results:
How to Run Accumulated Measurements
To run accumulated measurements:
1
Ensure that the instrument is set up correctly and that the error detector is
already in synchronization.
2
Click the Accumulation Setup menu item from the ED Setup submenu to check
the settings made for the data accumulation.
See “Error Accumulation - Procedures” on page 220 for details.
3
To begin accumulation, press the Start Accum button on the front panel.
This causes the following events:
If Prompt for File Name has been selected during setup, you will be asked to
enter a name for the log file before accumulation begins.
The data display in the Accumulated Results window is reset.
The recording of data to the measurement log file is reset.
How to Stop Accumulated Measurements
If you specified a single or repeat test, the test runs until the specified end condition
is reached. If you selected a manual test, it will run until you stop it manually.
Press the Stop Accum button on the front panel to stop the test immediately.
N O T E
Errors are only counted when the signal at the Gate In port is low.
Evaluating Results 7
Agilent J-BERT N4903B High-Performance Serial BERT 377