Eye Diagram - Procedures
This section lists out some of the generic processes to set up measurements, to
run them, and to create Eye Diagrams.
In this section, you will learn about:
“How to Prepare the Eye Diagram Measurement?” on page 320
“How to Initiate an Eye Diagram Measurement?” on page 321
“How to Change the Default Settings of an Eye Diagram?” on page 322
“How to Set Markers?” on page 325
“How to Run the Mask Measurement?” on page 326
“How to Customize Mask Files” on page 327
“How to Enter Notes?” on page 329
“How to Save and Recall” on page 329
“How to Export Fetched Data?” on page 330
How to Prepare the Eye Diagram Measurement?
An example to prepare an Eye Diagram measurement:
1
Disable the pattern generator outputs by pressing the 0V (Disable) button in
the PG Setup -> Data Output screen.
2
Connect the DUT to the pattern generator’s Data Out port and the error
detector’s Data In port. Refer to the Getting Started Guide.
3
Terminate all non-connected pattern generator output ports with 50 Ω.
4
Switch to the Pattern panel and click Pattern Select.
5
Select an appropriate pattern, for example a pure 2^15–1 PRBS, for both the
pattern generator and the error detector.
6
Specify the logic levels and the bit rate for the pattern generator setup. For
example, select ECL levels and a clock speed of 1250 MHz. This corresponds
to a clock period of 0.8 ns.