User guide

Low Level
Specifies the lower limit of the measurement voltage range. Enter the value in
mV. This value should be slightly lower than the lowest expected signal voltage.
High Level
Specifies the upper limit of the measurement voltage range. Enter the value in
mV. This value should be slightly higher than the highest expected signal
voltage.
Pass/Fail Tab
The Pass/Fail tab of the Properties dialog allows you to specify the criteria to
decide whether the DUT passes or fails the test. You can change pass/fail criteria
without rerunning the test. The software only uses the criteria to rate the results
of a measurement.
N O T E
The pass/fail criteria do not control measurement execution. The measurement run
will be completed even if the measurement fails one or more of the criteria.
You can set several pass/fail limits. For an explanation of the Eye Opening Pass/
Fail criteria, see “Measurement Parameters” on page 301.
6 Advanced Analysis
296 Agilent J-BERT N4903B High-Performance Serial BERT