User guide

Fast Total Jitter at BER
This enables the Fast Total Jitter measurement. Before enabling this
measurement, you need to know the BER floor of the device and to specify
a BER threshold that is above that floor. For details see “Explanation of the
Fast Total Jitter Measurement” on page 246.
Pass/Fail Tab
The Pass/Fail tab of the Properties dialog box allows you to specify the criteria
to decide whether the DUT passes or fails the test. You can change pass/fail
criteria without rerunning a test. The software only uses the criteria to rate the
results of a measurement.
N O T E
The pass/fail criteria do not control measurement execution. The measurement run
will be completed even if the measurement fails for one or more of the criteria.
You can set pass/fail limits individually for the output timing parameters and the
jitter measurement parameters:
For information on the DUT Output Pass/Fail criteria, see also “Output Timing
Measurement Parameters” on page 259.
For information on the Jitter Pass/Fail criteria, see also “Jitter Measurement
Parameters” on page 260.
6 Advanced Analysis
254 Agilent J-BERT N4903B High-Performance Serial BERT