The parameters can be changed to configure the output according to the DUT’s
termination requirements. Be aware that the changes are not applied immediately
but only when the output protection circuit button is clicked.
Output Protection on M8061A Output(s)
The output protection state on M8061A output(s) occurs when the amplifier
detects an overload condition which is caused by the externally connected circuitry
(DUT). In the output protection state, the Auto Range, Amplitude Range, Amplitude,
High, Low, Offset, Coupling, Termination Model, Termination Voltage and CMI state
are re-programmed to safe values.
The GUI shows an error message, listing the effected outputs and highlights the
Outputs On status indicator:
In the M8061A window, the parameters that currently do not match the data out
settings are highlighted in yellow.
The parameters can be changed to configure the output according to the DUT’s
termination requirements. The changes are applied only when the output
protection circuit button is clicked.
Setting up the Pattern Generator 4