Caution:
Take special precautions with MEDICAL ELECTRICAL EQUIPMENT regarding EMC. It must be
installed and put in service according to the EMC information provided above in International
Regulatory Statements of Conformity.
Caution:
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
Caution:
The use of accessories other than those specified may result in increased emissions or
decreased immunity of the device.
Warning:
The equipment should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, you must observe the equipment to verify normal operation in the
configuration in which it is to be used.
Declarations and Specifications
Guidance and manufacturer’s declaration – electromagnetic immunity
The ONA 1101GT is intended for use in the electromagnetic environment specified below. The customer or
the user of the ONA 1101GT should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic
environment - guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
<5 % Ut
(>95 % dip in Ut)
for 0.5 cycle
40 % Ut
(60 % dip in Ut)
for 5 cycles
<5 % Ut
(>95 % dip in Ut)
for 0.5 cycle
40 % Ut
(60 % dip in Ut)
for 5 cycles
Table continues…
December 2017 Open Networking Adapter 1101GT Installation Job Aid 6