User manual

WQ500 Series-EU-EN V1.2 10/13
6
ORPCalibration
CalibrationisnotnecessaryforORP.However,ORPshouldbeverifiedwithORPstandardsolution
tocheckwhethertheelectrodeisinproperworkingorderornot.
Conductivity,TDS,andSalinityCalibration
1. DiptheConductivitycellandthetemperatureprobeintothestandardsolution1413μS/cm.
2. Stirgentlyandwaituntilthereadingisstable.
3. Press and hold the Power button for 3 sec. to enter the calibration mode. The display will
showCALand1413μS/cmwillflash.
4.
When calibration is complete the display will stop flashing and indicate SA, then End. The
meterwillthenreturntothenormaloperatingmode.
Notes:
1. Calibratingwith12.88mS/cmstandardsolutionisbetterformeasuringwithhighconductivity
solutions.
2. TheiconCONDwilldisplayautomaticallyduringcalibrationmode.
3. Calibrationerrorindicatoriconwillappear(ErrinsteadofSA)ifcalibrationfails.
4. Ifthereadingisnot0μS/cmwhilethecellisinair(notdippedintoanysolution),calibratein
airtoobtainareading
of0μS/cm.
5. The Conductivity calibration points are: 0, 84μS/cm, 1413μS/cm, 12.88mS/cm and
80.0mS/cm.
DOCalibration
1. WiththeDOprobeattached,pressPowerbuttontoturnonmeterandthenpressMODEto
selectO2mode.
2. Wait10to30minutesfortheprobetopolarize.Thereadingshouldbeapprox.101.7%
(saturation)aftertheprobehascompletelypolarized
3. Removetheprobe.PressandholdPowerfor3sec.toentercalibrationmode.Thedisplaywill
showCALand101.7%willflash.
ThedisplaywillstopflashingandindicateSA,thenEndwhen
calibrationiscomplete.Themeterwillreturntothenormaloperatingmode.
4. Optional’zerooxygen’calibration:(improvesmeasurementaccuracyforveryloworveryhigh
DOmeasurements).Placetheprobeintoazerooxygencalibrationsolution,suchas
5%
sodiumsulfite,waitforstabilityandpressandholdPowertoentercalibration.Stabilityina
zerosolutionmaytakemanyminutes,dependingonprobehistory.
Note:
1. TheiconO2willdisplayautomaticallyduringcalibrationmode.
2. Calibrationerrorindicatoriconwillappear(ErrinsteadofSA)ifcalibrationfails.
3. Ifthe readingisnot0% whentheprobe isdisconnected,calibrateit inairwithoutprobe to
obtainareadingof0%.