User manual

WQ500 Series-EU-EN V1.2 10/13
7
pHMeasurement
1. PressMODEtochoosepHmode.
2. Aftercalibration,rinsethepHelectrodewithcleanwaterandwipeitdry.
3. Dip the electrode and the temperature probe into the sample solution that is going to be
measured.
4. Stirgentlyandwaituntilastablereadingcanbeobtained.
ORPMeasurement
1. InserttheORPelectrode,andpressMODEtochooseORPmode.
2. RinsetheORPelectrodewithcleanwaterandwipeitdry.
3. Diptheelectrodeintothesamplesolutionthatisgoingtobemeasured.
4. Stirgentlyandwaituntilastablereadingcanbeobtained.
Notes:
1. Thedisplaywillshow‐‐‐‐inanoverrangecondition.
2. Aftermeasurement,rinsetheelectrodewithcleanwater.
3. Replacethesoakingbottle.Thebottleshouldalways be filledwithsoakingsolution(4MKCL).
COND,TDS,SalinityMeasurement
1. InsertConductivitycell,andpressMODEtochooseCOND,TDSorSaltmode.
2. Aftercalibration,rinsetheconductivitycellwithcleanwaterandwipeitdry.
3. Dip the electrode and the temperature probe into the sample solution that is going to be
measured.
4. Stirgentlyandwait
untilastablereadingcanbeobtained.
Notes:
1. Thedisplaywillshow‐‐‐‐inanoverrangecondition.
2. ThemeterautorangestoμS/cmormS/cm,orppmorppt.
3. Aftermeasurement,rinsethecellwithcleanwaterandreplacetheprotectivecap.
4. Donot
touchorwipethesurfaceoftheinnerblackplateoftheconductivitycell.
DOMeasurement
1. RemovetheprotectioncapandturnonthepowerandpressMODEtochooseO2mode.Wait
10to30minutesfortheprobetopolarize.Thereadingshouldbeapprox.101.7%(saturation)
aftertheprobehascompletelypolarized.
2. SelectthedesiredunitsofmeasurebypressingMODEuntil
theproperunitsareshowninthe
display.
3. Placetheprobeinthesampletobemeasured.Stirtheprobeinthesampletoremoveany
trappedairbubblesfromthemembranesurface.
4. Allowthemetertimetosettletothefinalmeasurementvalue.