MembraneCapReplacement
IMPORTANTNOTE:OnlytheelectrolytesolutionsuppliedbyExtechwillworkcorrectlyforthis
probe.Ifathirdpartysolutionisused,theprobewillnotworknormally.
1. Donottouchthemembraneasskinoilswillinterferewiththeoxygenpermeabilityrateofthe
membrane.Replacethecapcarefully.
2. Itisrecommendedthattheproberemainattachedtothemeterduringthisreplacement
process.
3. Unscrewthecapfirmlyandcarefullyfromtheprobe.
4. RinsetheoldelectrolytesolutionfromtheCathodeandAnode.
5. UsethesuppliedPolishingStripstoclean,polish,shine,and/orremovescratches
fromthe
cathode.Besuretomoistentheclothbeforepolishingthecathode.Donotoverpolishthe
sensitivegoldcathode.
6. Setthenewreplacementmembranecaponaflatsurface.Leavethecapinthisposition
duringthereplacementprocess.
7. Fillthemembranecapwiththeelectrolytesolution
uptothebottomofthethreadsonthe
insideofthecap.
8. Tapthemembranecaptoreleaseandpreventairbubbleinelectrolytesolution.
9. Keepingthecapinafixedpositiononaflatsurface,carefullyinserttheprobeintothenew
capbyfirstdipping
andremovingtheprobeseveraltimesfromthecap.Witheachdip,push
theprobeprogressivelydeeperintothebondedcap.Finally,screwtheprobeslowlyontothe
capuntilfullytightened.Thedippingandremovaltechniqueminimizestheintroductionofair
bubblesintotheelectrolytesolution.Airbubblesinthe
electrolytecanaffectmeasurements.
10. Itisnormalthatexcesselectrolytesolutionwillleakfromthecapduringthisreplacement
sinceitminimizestheintroductionofairpockets.Cleanofftheexcesselectrolytebeforeuse.
1) MembraneCap
2) Cathode
3) Anode
4) DOProbe
CleaningtheDOProbe
WhentheDOreadingisunstableorincorrect,theprobeneedstobecleaned:
1. Unscrewthemembranecapfromtheprobe.
2. Placesandpaperonatablewiththecoarsesidefaceup,andaddsomewater.
3. Rubthecathodepartagainstthesandpaperabout10times.
4. Usethesandpapertocleantheanodepart.
5. Aftercleaningtheprobe,pleaserefillmembranecapwithnewelectrolyte.