User manual
WQ500 Series-EU-EN V1.2 10/13
9
AdvancedSettings
InpH,TDS,orDOmodepresstheupanddownarrowssimultaneouslyfor3secondstoenterthe
AdvancedSetupmode.
pHAdvancedSettings
1. InpHsettingmode,pressSTORE/RECALLbuttontosetthepHcalibrationtype.Usetheup
anddownarrowstoselectUSAorNIST.USAandNISTcalibrationpointsdiffer,refertothe
Calibrationsectionofthisguideforcomparisonofcalibrationpoints.
2. PressMODEtosavesettingandreturntonormaloperatingmode.
3. PressLOCK/MAXMINtoturnAUTOLOCKONorOFF.Usetheupanddownarrowstoselect
ONorOFF.InAutoLockmodethekeypadislockedandcannotbetampered.
4. PressMODEtosavesettingandreturntonormaloperatingmode.
TDSAdvancedSettings
1. PressSTORE/RECALLtosettheratiobetweenconductivityandTDS.
2. Usetheupanddownarrowstoadjusttheratiofrom0.5to1.0andthenpressMODEtosave
settingandreturntonormaloperatingmode.
DOAdvancedSettings
1. PressSTORE/RECALLtosetthesaltcompensation.Usetheupanddownarrowstoadjustthe
valuefrom0to50ppt.
2. PressMODEtosavesettingandreturntonormaloperatingmode.
3. PressLOCKtosetthealtitudecompensation.Usetheupanddownarrowstoadjustthe
valuefrom0to7000m(20000ft.).
4. PressMODEtosavesettingandreturntonormaloperatingmode.
Note:Inanyadvancedsettingmode,pressthepowerbuttontoresetallsettingsbacktofactory
setup(exceptpHcalibrationandAutolock)
PCInterface
ThisWQseriesmetercanbeconnectedtoaPCfordatatransferandotherremotefunctionality.
PleaserefertothePCInterface SoftwareInstructions suppliedontheincludedprogramCD‐ROM
and/ortheinstructionsprovidedintheHELPUtilityoftheSoftwareProgramitselfonceitis
opened.Pleasecontact
ourCustomerCareTechnicalSupportdepartmentforfurtherPC
interface‐relatedassistance.Checkthewebsitewww.extech.comsoftwaredownloadpagefor
thelatestversionofprogramandfortheoperatingsystemcompatibility.