User manual

2 TL625-EN-V1.0 4/12
Specifications
Attenuation 1:1000
Input Resistance 100MΩ
Input Capacitance X100:14.5pF to 17.5pF
Compensation Range 15pF to 35pF
System Bandwidth X100 DC to 100MHz
Maximum Working Input Voltage X100: <2000VDC + Peak AC
Cable Length 47” (120cm)
Weight 0.15lb (65g)
Operating Temperature 14F to 122F (-10C to 50C)
Storage Temperature -4F to 167F (-20C to 75C)
Humidity <85% RH
Maximum Working Voltage Derating Curve (VDC + Peak AC)
Low Frequency Probe Compensation
Before taking any measurements using the probe, first check the compensation and adjust it to match
the channel inputs. Most oscilloscopes have a square wave reference signal available at a terminal on
the front panel used to compensate the probe. Connect the probe to the signal source to display a
1kHz test signal on the oscilloscope.
Adjust the trimmer “L” until the signal displays a flat-top square wave.