11 SDL200-EU-EN-V1.2-7/11
Electrical Specifications
(ambient temperature 23
°
C ± 5
°
C)
Pt 100 ohm RTD Input
(0.00385 alpha coefficient meeting DIN IEC 751)
Resolution Range Accuracy
0.1
°
C
-199.9 to 850.0
°
C
± (0.4 % + 1
°
C)
0.1
°
F
-327.0 to 999.9
°
F
± (0.4 % + 1.8
°
F)
1
°
F
1000 to 1562
°
F
± (0.4 % + 2
°
F)
Thermocouple Input
(types K, J, T, E, R, and S)
The supplied thermocouple probes are useable in the range of -20 to +250°C (-4 to +482°F)
only. For measuring outside of this range, please obtain thermocouples rated accordingly.
Accuracy specified for meter only. Input probes add additional measurement error.
Generally RTD probes introduce a smaller error than thermocouples.
The above specifications are tested under an environmental RF Field Strength lower than 3
V/M and a frequency lower than 30 MHz
Copyright © 2011 Extech Instruments Corporation (a FLIR company)
All rights reserved including the right of reproduction in whole or in part in any form
Sensor Type Resolution Range Accuracy
-50.1 to -100.0
°
C
±(0.4 % + 1
°
C)
0.1
°
C
-50.0 to 999.9
°
C
±
(0.4 % + 0.5°
C)
1°C
1000 to 1300
°
C
±(0.4 % + 1
°
C)
-58.1 to -148.0
°
F
±(0.4 % + 1.8
°
F)
0.1
°
F
-58.0 to 999.9
°
F
±(0.4 % + 1
°
F)
Type K
F
1000 to 2372
°
F
±(0.4 % + 2
°
F)
-50.1 to -100.0
°
C
±
(0.4 % +
C
)
0.1
°
C
-50.0 to 999.9
°
C
±(0.4 % + 0.5
°
C)
C
1000 to 1150
°
C
±(0.4 % + 1
°
C)
-58.1 to -148.0
°
F
±(0.4 % + 1.8
°
F)
0.1
°
F
-58.0 to 999.9
°
F
±
(0.4 % +
F
)
Type J
1
°
F
1000 to 2102
°
F
±(0.4 % + 2
°
F)
-50.1 to -100.0
°
C
±(0.4 % + 1
°
C)
0.1
°
C
-50.0 to 400.0
°
C
±(0.4 % + 0.5
°
C)
-58.1 to -148.0
°
F
±(0.4 % + 1.8
°
F)
Type T
0.1
°
F
-58.0 to 752.0
°
F
±
(0.4 % +
F
)
-50.1 to -100.0
°
C
±(0.4 % + 1
°
C)
0.1
°
C
-50.0 to 900.0
°
C
±(0.4 % + 0.5
°
C)
-58.1 to -148.0
°
F
±(0.4 % + 1.8
°
F)
0.1
°
F
-58.0 to 999.9
°
F
±
(0.4 % +
F
)
Type E
1
°
F
1000 to 1652
°
F
±(0.4 % + 2
°
F)
0 to 600
°
C
±(0.5 % + 3
°
C)
1
°
C
601 to 1700
°
C
±(0.5 % + 3
°
C)
32 to 1112
°
F
±(0.5 % + 5
°
F)
Type R
F
1113 to 3092°F
±
(0.5 % +
F
)
0 to 600
°
C
±(0.5 % + 3
°
C)
1
°
C
601 to 1500°C
±(0.5 % + 3
°
C)
32 to 1112
°
F
±(0.5 % + 5
°
F)
Type S
F
1113 to 2732
°
F
±
(0.5 % +
F
)
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com