User Manual
10 SDL200-EU-EN-V1.2-7/11
Battery Replacement and Disposal
When the low battery icon appears on the LCD, the batteries must be replaced. Several hours of
accurate readings are still possible in this condition; however batteries should be replaced as soon
as possible:
• Remove the two (2) Phillips screws from the rear of the meter (directly above the top of the tilt
stand).
• Remove and safely place the battery compartment and screws where they will not be damaged
or lost.
• Replace the six (6) 1.5V ‘AA’ batteries observing polarity.
• Replace the battery compartment cover with the two (2) Phillips screws.
All EU users are legally bound by the battery ordinance to return all used batteries to
collection points in your community or wherever batteries / accumulators are sold!
Disposal in the household garbage is prohibited!
Specifications
General Specifications
Display Backlit LCD; LCD size: 52 x 38mm (2 x 1.5”)
Status indicators Over-range (----) and low battery
Measurement Channels T1, T2, T3, T4, and T1-T2
Sensor types Thermocouple types: K, J, T, E, R, and S
RTD type: platinum 100 ohm (Complies with 0.00385 alpha
coefficient meeting DIN IEC 751)
Measurement Units °C / °F
Offset Adjustment Available for one (1) thermocouple and two (2) RTD inputs
Input Linearity Sensor curves are stored in meter’s CPU
Datalogger Sampling Rate AUTO LOGGING: 1, 2, 5, 10, 30, 60, 120, 300, 600, 1800,
3600 seconds. Note that a one (1) second sampling rate
can cause some data loss on slower computers
MANUAL LOGGING: Set the sampling rate to ‘0’
Memory Card SD memory card; 1G to 16GB size
Temperature Compensation Auto temperature compensation for all thermocouple types
Display update rate Approx. 1 second.
Data Output RS-232 / USB PC computer interface; for use with optional
software kit part number 407001A
Operating Temperature 0 to 50°C (32 to 122°F)
Operating Humidity 85% R.H. max.
Auto Power OFF After 10 minutes of inactivity (can be disabled)
Power Supply Six (6) 1.5 VDC batteries (optional 9V AC adaptor)
Power Consumption Normal operation (backlight & datalogger OFF): approx. 9.7mA dc
With backlight OFF and datalogging ON: approx. 30mA dc
With backlight ON add approx. 17mA dc
Weight 347g (0.76 lbs.) meter only
Dimensions Main instrument: 182 x 73 x 47.5mm (7.1 x 2.9 x 1.9”)