User Manual

SD200-EU-EN-V1.2-7/10
6
Advanced Settings
The SET function is used to:
Format the SD memory card
Set the date and time
Set the sampling time
Set the beeper sound ON/OFF
Set the SD card Decimal character
Select the Temperature units
Set the RS232 data output ON/OFF
1. Press and Hold the SET button for >2 seconds to enter the setting mode. The first
function (Sd F) will appear in the display. Press the SET button to step through the
seven functions. Use the and buttons to adjust the selected function. Use the
“LOGGER” button to step through fields within a function. In the SET mode, if no
button is pressed within 5 seconds the logger will revert back to the standard mode.
2. Sd F – Format the SD card. Press the button to select yES or no. For yES, press
the Enter button to format the card and erase all existing data.
3. dAtE – Set the date and time. Press the or buttons to adjust the selected
(blinking) field. Press the Enter button to store the value and to step through the
various fields.\
4. SP-t – Set the sample rate. Press the button to select the desired sample rate and
press Enter to store the selection. The selections are: 5, 10, 30, 60, 120, 300, 600
seconds and AUTO. In AUTO, the data will be stored every time there is a
temperature change of >1 degree.
5. bEEP - Set the beeper ON or OFF. Press the button to select ON or OFF and
press Enter to store the selection.
6. dEC - Set the SD card Decimal character. Press the button to select USA
(decimal) or Euro (comma) and press Enter to store the selection.
7. t-CF - Set the Temp. unit to °F or °C
8. rS232 - Set the RS232 data output ON/OFF. Press the button to select ON or off
and press Enter to store the selection
9. ESC – Exit the setting mode. Press the SET button to return to normal operation.
System RESET
If a condition appears where the CPU does not respond to keystrokes or the meter seems
frozen, press the RESET button on the side of the datalogger (use a paper clip or similar
pointed object) to return the meter to a working state.
RS232 Interface
A serial output is provided to link the meter to a PC’s serial port. Call Technical Assistance
for more information on this interface.