User manual
PQ3350 V5.2 11/10 35
For example, ±0.8W becomes ± 0.8W * CT ratio
Range of CT (Current Transformer) Ratio: 1 to 600
AC Apparent Power (VA, from 0.000VA to 9999 KVA):
VA = V r.m.s. x A r.m.s
AC Reactive Power (VAR, from 0.000 VAR to 9999 KVAR):
VAR =
√
(VA
2
– W
2
)
AC Active Energy (mWH, WH, or KWH, from 0 mWH to 999,999 KWH)
WH = W * Time (in hours)
AC Current
(50 or 60 Hz, Auto Range, True RMS, Crest Factor < 4, CT=1)
Model PQ3350+PQ3110 (Overload Protection AC 200A)
Range
Resolution
Accuracy of Readings
4
0.04
–
1
A
0.001 A
±0.5% ±
0.0
5
A
0.4
–
1
0 A
0.01 A
±0.5% ±
0.0
5
A
4
–
100
A
0.1 A
±1.0% ±
0.
5
A
Model PQ3350+PQ3120 (Overload Protection AC 2000A)
Range
Resolution
Accuracy of Readings
5
10.00A
0.001A/
0.01A
–
4A
-
100.0A
0.01A/
0.1A
±0.5% ±
0.
5
A
40A
–
1000.0 A
0.1A/
1 A
±0.5% ± 5
A
Model PQ3350+PQ3220 (Overload Protection AC 3000A)
Range
Resolution
Accuracy of Readings
6
0
–
300.
0A
0.1A
±
1
%
of range
300.0
–
3000A
0.1A / 1A
±
1
%
of range
Model PQ3350+PQ3210 (Overload Protection AC 1200A)
Range
Resolution
Accuracy of Readings
6
0
–
120.0A
0.1A
±
1
%
of range
120.0
–
1200A
0.1A / 1A
±
1
%
of range
4, 5, 6
For CT
≠
≠≠
≠
1, the accuracy in percentage is the same (±0.5%). But the
additional digits should be multiplied by the CT ratio.
For example, ±0.5A becomes ±0.5A * CT ratio