Getting Started
Remove the cap from the bottom of the ExStik
TM
to expose the electrode bulb and
reference junction
Before first use or after storage, soak the electrode (with its cap removed) in a pH 4
solution for about 10 minutes (PH100 only)
White KCL crystals may be present in the cap. These crystals will dissolve in the soak or
they can be simply rinsed with tap water
Always calibrate close to the expected measurement value
Twist off the battery compartment cap to access the battery tab; remove this plastic tab
before first use
To preserve Electrode life keep the sponge in the protective cap soaked with a pH 4
solution (PH100), deionized, distilled water (RE300) when not in use (store vertically)
Specifications
Display Multifunction LCD with Bargraph
Operating conditions 32 to 122
o
F (0 to 50
o
C) / < 80% RH
pH Range and Accuracy 0.00 to 14.99 / ± 0.01pH typical
Temp. Compensation Automatic from 32 to 194
o
F (0 to 90
o
C)
Compensation Accuracy ± 0.9
o
F (0.5
o
C)
ORP Range / Accuracy ± 999mV / ± 4mV
Temperature Range 23 to 194
o
F (-5 to 90
o
C)
Temperature Resolution 0.1
o
up to 99.9 then 1
o
thereafter
Temperature Accuracy ± 1.8
o
F / 1
o
C [from 23 to 122
o
F (-5 to 50
o
C)]
± 5.4
o
F / 3
o
C [from 122 to 194
o
F (50 to 90
o
C)]
Measurement storage 15 tagged (numbered) readings
Power Four (4) SR-44 button batteries (See note on page 6)
Low battery indication ‘BAT’ appears on the LCD
Auto power off After 10 minutes of inactivity
Battery Life 200 to 400 tests, assuming test time of < 10 minutes and use of
SR44 silver oxide batteries. Use of LR44 alkaline will reduce
battery life.