Moisture Alarm Set Procedure:
1. With MOIST displayed, simultaneously press the MOIST and MODE buttons.
2. The “HIGH” icon will appear on the display.
3. Press the ▲or ▼button to set the high limit desired.
4. Press the STORE/ALARM SET button to save the value and proceed to the LOW set
value.
5. With the “LOW” icon in the display, Press the ▲or ▼button to set the low limit
desired.
6. Press the STORE/ALARM SET button to save the value and to return to the normal
mode.
7. If the moisture measurement is lower than the LOW alarm setting, the meter will beep
once every second
8. If the moisture measurement is higher than the HIGH alarm setting, the meter will
beep continuously.
Data Memory Storage
Storing Readings:
1. With the data to be saved in the display, press the STORE button for 2 seconds until
the unit beeps. The data will be saved to the memory location indicated. The numeric
display above the MEM icon will then advance to the next location.
2. When the 20 memory locations are full, the unit will overwrite old saved readings
starting with memory location 01.
Recalling Stored Readings:
1. Simultaneously press and release the ▲and ▼ buttons to display stored readings.
The numeric display above the MEM icon will flash and the data stored in that location
will be displayed.
2. Press the ▲or ▼button to scroll through the memory locations.
3. To return to normal operating mode, press the STORE button.
Clearing Stored Data:
1. To clear stored data, press and hold the ▲and STORE buttons simultaneously until
CLR appears on the display.
Auto Power Off
The meter will enter a sleep mode after 30 minutes of inactivity. The meter will emit a
warning beep 15 seconds before shutting down.
To disable the APO feature, press the MODE button when turning the meter ON. The
“APO” icon will not appear, indicating it is disabled.