Relative / % Function 
The REL/% mode allows for % deviation measurements from a stored reference value. 
1.  Setup the meter with the parameters required for the test to be performed. 
2.  Insert the reference component into the test fixture and wait for a stable reading. 
3.  Press the REL/% button to store the value. The  icon will appear in the display. 
4.  For all subsequent measurements, readings on the lower display will indicate the % 
difference between the currently measured component and the stored value. 
5.  Press and hold the REL/% button for >2 seconds to exit the mode. 
Sorting 
The sorting mode is used to select components within a % limit of a reference value. 
1.  Insert the reference component and set the test parameters. “LCR” auto-parameter is not 
allowed in the sorting mode. 
2.  With the desired reading in the display, press the SORT button to establish the reference 
value. The main display will indicate “PASS” and the lower display will indicate the value of 
the component. The default sorting % is +/- 1%. 
Note: If the reference value is above 2000 counts or below 200 counts, sort will not work. 
3.  Press the SETUP button in sequence to change the Range, Reference Value, and the % 
Tolerance. 
a.  Range: With the “RANGE” icon flashing, press the  or  button to change the range. 
Press the ENTER button to store the setting and proceed to the Value setting. 
b.  Value Setting: Press the  or  button to select the flashing digit for adjustment. Press 
the  or  button to adjust the value of the digit. Press the ENTER button to store the 
setting and proceed to the Tolerance setting. 
c.  Tolerance setting: Press the  or  button to step through the available tolerance 
selections: 
± 0.25% 
± 0.5% 
± 1% 
± 2% 
± 5% 
± 10% 
± 20% 
+80% -20% 
Press the ENTER button to store the Tolerance setting. 
4.  Press the SORT button for > 2 seconds to exit the Sorting mode. 