User Manual
4 LCR200-EU-EN-V1.2 7/11 
Operation 
Setup 
1. Press the 
 Power button to turn the meter on. 
2.  The meter will turn on in the default mode: AUTO LCR and 1kHz 
3.  APO will be active with auto shutoff every 5 minutes unless an AC adaptor is used. 
LCR primary parameter selection 
1.  The meter will turn on with auto parameter selection enabled and with the ”APO” (auto power 
off), “Auto” (autoranging) and “LCR” (auto-parameter) icons appearing at the top of the 
display. 
2.  To manually select the parameter, press the L/C/R button to step through and select the 
parameter needed. Each button press will sequentially display: 
Auto-LCR  Auto Range  Auto Parameter 
Auto-L Auto Range  Inductance 
Auto-C Auto Range  Capacitance 
Auto-R Auto Range  Resistance 
DCR  DC Resistance 
3.  The value of the primary parameter will be displayed in the upper display and the secondary 
parameter in the lower display. 
D/Q/θ secondary parameter selection 
1.  The meter will turn on with auto parameter selection “LCR” enabled. .The primary and 
secondary parameter will be automatically selected based on the value of the measured 
impedance. 
2.  To manually select the secondary display, first select the primary display. 
3.  Press the D/Q/ button to select the secondary parameter: 
L  D, Q, ESR(RP) or  
C  D, Q, ESR(RP) or  
R none 
DCR none 
4.  The value of the secondary parameter will be displayed in the lower display 
Series or Parallel 
1.  When the L/C/R function mode is selected and ‘AUTO” is enabled, the default measurement 
in serial or parallel mode is auto selected. A parallel equivalent circuit (Lp, Cp or Rp) will be 
selected if the impedance is greater than 10k. A series equivalent circuit ( Ls, Cs or Rs) will 
be selected if the impedance is less than 10k. 
2.  Press the SER/PAR button as needed to change the default selection. 
Frequency 
Press the FREQ button to change the test frequency. The selections are: 100Hz, 120Hz, 1kHz, 
10kHz and 100kHz. 
Data Hold 
Press the HOLD button to freeze the reading in the display. Push the HOLD key again to cancel 
the hold mode and return to normal operation. 