User manual

EXTECHINSTRUMENTS
7IR300UVenGB_V1.01/16
UVLeakDetector
Injectasmallamountofflorescentdyeintothesystemundertest.Thedye/systemmixture
(refrigerant,forexample)willescapeandaccumulateatallleaksites.TurntheblueUVLEDson
intheSettingmodeandscanthesystem.Theresidueofthedyeattheleakcanbeeasily
observedundertheUVlight.DyesareavailableatmostHVACsupplyoutlets.
LocatingHotorColdSpots
Todetectahotorcoldspot,aimthethermometerataregionbeyond
thetargetandthenscantheentireregioninaslow,up/downmotion.
DistancetoSpotRatio(FieldofView)
Themeter’s fieldof view is 12:1(distance tospotratio). For example,if themeteris60cm (24
inches)fromthetarget(spot),thediameterofthetargetmustbeatleast5cm(2inches).Other
distancesareshownbelowinthefieldofviewdiagram.
Note that measurements should normally be made further than 60cm (24 inches) from the
target.Themetercanmeasurefromfurtherdistancesbutthemeasurementmaybeaffectedby
externalsourcesoflight. In addition,thespotsize may besolarge that it encompassessurface
areasnotintendedtobemeasured.
2.5cm@30cm 5cm@60cm 10cm@120cm
1”@12” 2”@24” 4”@48”
Itis necessary to ensure thatthe sizeof thetarget is largerthan thespot size.The smallerthe
target,thecloserthedistanceshouldbe.