User manual
EX570-EU_ENG V2.3 3/09 15
Specifications
Function Range Resolution Accuracy
400 mV 0.01 mV
4V 0.0001V
40V 0.001V
400V 0.01V
DC Voltage
1000V 0.1V
±(0.06% reading + 2 digits)
40 Hz to 65 Hz 66 Hz to 1000 Hz
400 mV 0.01 mV
4 V 0.0001 V
40 V 0.001 V
400 V 0.01 V
1000 V 0.1 V
±(0.5% rdg + 3 digits) ±(1% rdg + 3 digits)
AC Voltage
All AC voltage ranges are specified from 5% of range to 100% of range
400μA 0.01 μA
4000μA 0.1 μA
40mA 0.001 mA
±(0.7% reading + 3 digits)
400mA 0.01 mA
±(1.0% reading + 3 digits)
10A 0.00 A
±(1.5% reading + 3 digits)
DC Current
(20 A: 30 sec max with reduced accuracy)
40 Hz to 65 Hz 66 Hz to 1000 Hz
400μA 0.01 μA
4000μA 0.1 μA
40mA 0.001 mA
400mA 0.01 mA
±(1.0% rdg + 3 digits) ±(1.5% rdg + 3 digits)
10A 0.001 A
±(2.0% reading + 3
digits)
±(2.0% reading + 3
digits)
AC Current
(20A: 30 sec max with reduced accuracy)
NOTE: Accuracy is stated at 65
o
F to 83
o
F (18
o
C to 28
o
C) and less than 75% RH.