User manual

EX503-EU_ENG V2.1 11/07 9
CAPACITANCE MEASUREMENTS
WARNING: To avoid electric shock, disconnect power to the unit under test and
discharge all capacitors before taking any capacitance measurements. Remove the
batteries and unplug the line cords.
1. Set the rotary function switch to the CAP position.
2. Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive V jack.
3. Press the MODE button to indicate “nF” on the display.
4. Touch the test leads to the capacitor to be tested.
5. The test may take up to 3 minutes or more for large capacitors to
charge. Wait until the readings settle before ending the test.
6. Read the capacitance value in the display
FREQUENCY/DUTY CYCLE MEASUREMENTS (ELECTRONIC)
1. Set the rotary function switch to the “Hz %” position.
2. Press the Hz/% button to indicate “Hz” in the display.
3. Insert the black lead banana plug into the negative COM jack and the
red test lead banana plug into the positive Hz jack.
4. Touch the test probe tips to the circuit under test.
5. Read the frequency on the display.
6. Press the Hz/% button again to indicate “%” on the display.
7. Read the % of duty cycle on the display.
FREQUENCY SENSITIVITY (ELECTRICAL)
The frequency sensitivity is range dependent when the Hz function is selected while in the
voltage or current measuring mode. Below are typical sensitivities for the “electrical”
measurement modes.
Range
( DC/AC )
Sensitivity Frequency width
4V
1.5V rms
5Hz~10kHz
10V rms
5Hz~20kHz
40V, 400V
20V rms
5Hz~200kHz
1000V/1000V
420V rms
50Hz~1kHz
400mA
45mA rms
5Hz~5kHz
10A
4A rms
5Hz~1kHz