User manual
AN200-EU-ENG V2.1 3/08
7
Air Flow Measurements (CMM / CFM)
1. Turn on the meter using the ON/OFF button
2. Press the UNITS button to select the desired air
flow units: CMM (cubic meters per minute) or
CFM (cubic feet per minute). NOTE: At power up
the meter will display the last unit of measure
previously entered.
3. To begin entering the area in m
2
or ft
2
, press and
hold the AREA button until it beeps twice. The
leftmost digit of the bottom display will begin to
flash.
4. Use the ▲ (UP) button to change the flashing
digit
Use the ◄ (LEFT) button to move the decimal
Use ► (RIGHT) button to select the other digits.
After all of the digits are entered, press and hold
the AREA button (until meter beeps twice) to
save the area into memory and return to CFM or CMM measuring mode.
5. Place the sensor in the air stream. Ensure that the air enters the vane as indicated by
the arrow sticker placed inside the vane. Refer to the diagram.
The meter has 16 memory locations (8 for CFM and 8 for CMM) that can be used to storel
commonly used area sizes that you can recall at anytime.
1. Press the AREA button until meter beeps twice. A memory location number will
appear in the top right of the display indicating the memory location.
2. Push the AREA button to scroll thru and select the desired location. Once you have
selected the desired memory location enter your dimension
Use the ▲ (UP) button to change the flashing digit
Use the ◄ (LEFT) button to move the decimal
Use ► (RIGHT) button to select the other digits. After all of the digits are entered,
press and hold the AREA button (until it beeps twice) to save the area into memory
and return to CFM or CMM measuring mode.
To select and use a previously stored dimension, press and hold the AREA button until it
beeps twice.
Press AREA to scroll thru the 8 memory locations. Press and hold the AREA button until it
beeps twice. to return to CFM or CMM measuring mode.
Air Flow Measurements (Up to 20 Point averaging)
1. To enter 20 Point Averaging Mode, press and hold the AVG button until it beeps twice.
The AVG icon will be displayed.
2. Take a measurement and press the AVG button. A single beep will sound and the
HOLD icon will appear in the display.
3. The average reading will be displayed and number of readings measured will appear in
the upper right hand corner of the display. After 5 seconds, the display will return to the
current reading. (IMPORTANT: Please note that the average readings are only held for
5 seconds and cannot be recalled.)
4. Repeat steps 2 - 3 until all desired points have been measured.
5. Press the AREA button to clear the multipoint averaging memory.
6. To return to standard airflow measuring mode press and hold AVG button until meter
beeps twice.
airflow
Arrow
Side view of Vane