5.4Auto1MemoryStorage
MemoryStore
Continuously store the current measured data as 1 set, at preset record interval times.
1. Power on the meter.
2. Press the MENU button to enter the menu list screen.
3. Use the
+
/
-
buttons to select “B)
.
Storage” then press the button.
4. Use the
+
/
-
buttons to select “FLASH” or “SDCard” then press the button.
Selecting FLASH will store the data in the internal memory.
Selecting SD Card will store the data on the SD CARD.
5. Use the
+
/
-
buttons to select “C)
.
St or e Mode” then press the button.
6. Use the
+
/
-
buttons to select “AUTO1” then press the button.
7. Use the
+
/
-
buttons to select “IntervalTime” then press the button.
8. Use the
+
/
-
buttons and the button to set the desired record interval time from 1
second to 23h59m59s, then press the button.
9. Use the
+
/
-
buttons to select “A)
.
Measurementtime” then press the button.
10. Use the
+
/
-
and button to set the desired measurement time from 1 second to
23h59m59s, then press the button.
Note:Measurement Time must be greater than or equal to the set Interval Time
11. Set the other measurement parameters (trigger mode, sub channel etc.) as required, using
the MENU button to select setting items.
12. Press the MENU button to return to the measurement screen.
13. Press the STORE button to start the measurement. If a trigger has been set, measurement
will start when the trigger conditions are met.
Important:During recording, only the START/STOP , PAUSE/CONT , , STORE , and M
buttons are operative. Other keys such as the FREQ WEIGHT TIME and LEVEL button
have no effect. Be sure to complete all settings before starting the store process.
If the trigger mode is “OFF”, storing of measurement data will stop when the end of the
measurement time is reached.