5.StoringDatatoMemory
The meter can store measurement data in the internal memory or on a micro SD card. There are three
different ways of storing data, as listed below.
Manual:
In this mode, the operator stores the measured sound level data and processed value data into the
memory manually.
Press the R record button to cause the current sound level and processed values derived from
the measurement, as well as the measurement parameters and time of recording to be recorded
in a single measurement.
Memory capacity: max. 250 data sets
Manually stored data can be recalled by the meter.
Auto1:
Soundlevelmetermode
Continuous store of main channel and sub channel all‐pass values as 1 set, at preset record
interval times.
OctaveAnalyzermode
Continuous store of main channel sound level Lp values for each band and all‐pass values, sub
channel all‐pass values as 1 set at preset record interval time.
Auto2:
Soundlevelmetermode
Continuous store of main channel and sub channel all‐pass values as 1 set at preset
measurement integrate time.
OctaveAnalyzermode
Continuous store of main channel sound level Lp values for each band and all‐pass values, sub
channel all‐pass values as 1 set at preset measurement integrate time.
Automatically recorded data can be retrieved using the supplied software.
5.1ManualMemoryStorage
MemoryStore:
Press the store key to store the current sound level and processing values. Only the sound level is
stored when the R button is pressed. There will be no processing values displayed when
the meter is powered on.
1. Power on the meter.
2. Press the MENU button to enter the menu list screen.
3. Use the
+
/
-
buttons to select “B)
.
Storage” then press the key.
4. Use the
+
/
-
buttons to select “FLASH” or “SDCard” then press the key.