User manual

407790A-en-US_v2.0 8/16
21
3.7TriggerModeSetupandOperation
Ameasurementcanbeinitiatedbyoneofthreetriggers:timetrigger(timecontrolled
triggering),leveltrigger(soundlevelcontrolledtriggering),andexternaltrigger(triggeringbyan
externalsignal).
Leveltrigger1:Measurementstartswhenthetriggerlevelisexceededandendsafterthe
measurementtime.
Thetriggerlevelisvalidatanytime.
Settingitems:
TriggerSlopeType:+,‐
TriggerLevel:25to130dB,1dBsteps.
TriggerBand:MAINAP/31.5Hz/63Hz/125Hz….8kHz(1/3octavebands)
Leveltrigger2:Singlemeasurementismadewhentriggerlevelisexceeded.
Thetriggerlevelisvalidatanytime.
Settingitems:
TriggerLevel:25to130dB,1dBsteps.
TriggerBand:MAINAP/31.5Hz/63Hz/125Hz…8kHz(1/3octavebands).
Timetrigger:Thisisareoccurringtriggertimerthatstartsameasurementafteratriggertimer.
Themeasurementendsafterthemeasurementtime.
Thetriggertimerisvalidatalltimes.
Settingitems:
TriggerTime:h:m:s
Externaltrigger:AMeasurementstartsatfallingedgeofalogiclevel(0to5Vlogic)signalatthe
externaltriggerinput.APulsewidthof1msormoreisrequired.
Theexternaltriggerisvalidatalltimes.
TriggerOperation‐example
1. SoundLevel
2. Time
3. Slope+Trigger
4. Slope‐Trigger
5. TriggerLevel
1
2
34
5