Calibration
CalibratingtheIntegratingSoundLevelDataloggerrequiresanacousticalcalibratorsuchasthe
ExtechModel407766or407744.
1. Configurethemeterasfollows:
Display:SPL(dBA)
Timeweightingmode:FAST
2. Insertthemicrophoneintotheopeningoftheacousticalcalibrator.
3. SwitchthecalibratorON.
4. Adjust
themeter’scalibrationpotentiometerasshowninthe
accompanyingdiagramuntilthemeter’sdisplaymatchesthe
acousticalcalibratoroutputsignal(typically94or114dB).
5. SwitchthecalibratorOFF.
6. Removethecalibratorfromthemicrophoneveryslowlytoavoid
damagingthemicrophone.
7. Incaseswhereadjustingthepotentiometer
doesnotaffectthe
displayedreading,orwheretheadjustmentcannotproducethe
appropriatedisplayedreading,pleasereturntheunitforservice.