User’s Manual Dual Laser Video IR Video Thermometer MODEL VIR50
Introduction Congratulations on your purchase of the Video IR Thermometer. This meter is capable of making non-contact (infrared) and contact (thermocouple) temperature measurements. The built-in camera offers still image and video capturing functionality for documenting measurement locations. The meter includes a micro SD card slot for offloading still images and video.
Meter Description 1. LCD Display 1 2. Push-buttons (UP/Picture, ESC, and DOWN/VIDEO) 3. Measurement Trigger 5 4. Retractable Lens Cover 2 5. Camera, Lasers, and IR sensor 4 6. Compartment for Type-K jack, USB jack, Reset button 7.
Operation Power ON and OFF 1. The instrument is powered by one (1) rechargeable battery (3.7V). The battery is housed in the compartment in the meter’s handle. 2. The supplied battery charger connects to the USB port on the instrument (in the compartment above the measurement trigger). 3. Once the battery is charged, Press and Hold the instrument ON. 4. Press and Hold the button for 5 seconds to switch the button to switch the unit OFF. 5.
Video + IR mode In the Video + IR Mode the meter can measure IR temperature, air temperature, humidity, dew-point temperature, and wet bulb temperature, along with full use of the video camera capabilities. The IR Maximum (MAX) Temperature, IR Minimum (MIN) Temperature, IR Differential (DIF) Temperature, and the IR Average (AVG) Temperature can also be displayed. Press and hold trigger to measure the temperature. Capturing Still Images (Photographs) button to take a photograph.
IR-only mode In the IR-only Mode the meter can measure IR temperature, air temperature, air humidity, dew-point temperature, and wet bulb temperature but without the video camera capabilities. The IR Maximum (MAX) Temperature, IR Minimum (MIN) Temperature, IR Differential (DIF) Temperature, and the IR Average (AVG) Temperature can also be displayed. Press and hold trigger to measure the temperature.
Datalogging Mode The datalogger automatically records readings to memory at a programmed interval. In the Datalogging mode, the high/low alarm values, low alarm value, datalogging interval sampling rate (time), can be configured. Once configured, press the trigger to start logging. The unit will automatically record data at the specified sampling time. Press the ESC buttom to exit the DATALOG mode, the meter will automatically save the data to a file.
System Tools Settings From the Video + IR, IR-only, or Dew Point modes, press the ENTER button to access the soft-key screen. The System Tools is the icon on the left. Use the arrow buttons to scroll over to the System Tools icon if it is not already highlighted. Press the ENTER button again when the System Tools icon is highlighted to access the System Tools settings mode. From the datalogger mode scroll down to the Measure Set parameter and press ENTER to access to the System Tools setting menu.
Setting the Language Use the ▲ and ▼ buttons to select the desired language, and press the ENTER button to confirm. Setting the Date Use the ▲ and ▼ buttons to select the desired format for the date, and press the ENTER button to confirm. Once the format has been chosen, use the ▲ and ▼ buttons to select the day, month, and year. Setting the Time Use the ▲ and ▼ buttons to select the desired format for the time, and press the ENTER button to confirm.
Setting the temperature units (C/F) Use the ▲ and ▼ buttons to select the desired unit of measure, press the ESC button to exit and save. Memory Use the ▲ and ▼ buttons to select either FLASH or SD CARD, and press the ENTER button to confirm. Then select either WORKING SPACE or FORMAT, and press the ENTER button to confirm. If FORMAT was pressed, click YES or NO to confirm. 10 VIR50-EU-EN v2.
Setting the Buzzer ON or OFF Use the ▲ and ▼ buttons to select either ON or OFF, and press the ENTER button to confirm. Setting the Auto Power off time Use the ▲ and ▼ buttons to disable or select an auto-power-off time, press the ESC button to exit and save. Setting the Screen Timeout Use the ▲ and ▼ buttons to disable or select a screen auto-off time, press the ESC button to exit and save. 11 VIR50-EU-EN v2.
Selecting the Memory Type If there is no SD card present in the VIR50, then only a single blue icon will be displayed on the LCD screen in the top left corner. If there is an SD card present, but the memory is being used by the device, then there will be two icons, a chip (blue i.e.
Measurement Settings From the Video + IR, IR-only, or Dew Point modes, press the ENTER button to access the soft-key screen. The Measurement Settings mode is represented by the middle icon. Use the arrow buttons to scroll over to the middle icon (if it is not already highlighted.) Press the ENTER button again when the middle icon is highlighted to access the Measurement Settings mode as shown below.
Emissivity setting Use the arrow buttons to highlight an emissivity setting and then press ENTER to choose it. Select emissivity by description (concrete, glass, etc.) or select ε=0.94 to set the emissivity manually using the arrow buttons. Press the ESC button to exit and save. Alarm High Use the ▲ and ▼ buttons enable or disable the alarm. If the high alarm is enabled, press the ENTER button to adjust its value; use the ▲ and ▼ buttons to adjust the value.
Alarm Low Use the ▲ and ▼ buttons enable or disable the alarm. If the low alarm is enabled, press the ENTER button to adjust its value; use the ▲ and ▼ buttons to adjust the value. Press the ENTER button to confirm and then press the ESC button to exit and save. Laser Press the ENTER button to enable or disable the laser. Auto Mode (Measure Lock) Press the ENTER button to enable or disable auto mode. Max/Min Press the ENTER button to set the MAX/MIN mode ON or OFF.
Type k Press the ENTER button to enable or disable the Type k contact thermocouple input.. NOTE: If a Type-k probe is inserted into the meter’s thermocouple jack (in the compartment over the measurement trigger), ‘Enable’ will be selected by default. Select the Disable setting to prohibit the LCD from displaying Type-k temperature. Color Use the ▲ and ▼ buttons choose a color. Press the ESC button to exit and save.
Item Description Picture Display saved images Video Playback saved videos Logs Display datalogger data 17 VIR50-EU-EN v2.
Transfer files to a PC Connect the USB cable from the thermometer to a USB port on the pc or insert the microSD card into the SD port on the pc.. If the USB cable is used, the USB symbol will appear in the display and the PC will recognize the unit as a removable drive. Open the drive to view the three folders: LOGS: Text files, use the *.txt files PICTURE: Pictures, use the *.jpg files VIDEO: Videos, use the *.
IR Measurement Notes 1. The object under test should be larger than the spot (target) size calculated by the field of view diagram (printed on the side of the meter and in this guide). 2. Before measuring, be sure to clean surfaces that are covered with frost, oil, grime, etc. 3. If an object's surface is highly reflective, apply masking tape or flat black paint to the surface before measuring. Allow time for the paint or tape to adjust to the temperature of the surface it is covering. 4.
Emissivity and IR Measurement Theory IR Thermometers measure the surface temperature of an object. The thermometer’s optics sense emitted, reflected, and transmitted energy. The thermometer’s electronics translate the information into a temperature reading which is then displayed on the LCD. The amount of IR energy emitted by an object is proportional to an object's temperature and its ability to emit energy.
Specifications Infrared Thermometer Specifications o o Range -58 to 3992 F (-50 to 2200 C) Resolution 0.1 C/F < 1000; 1 C/F > 1000 Accuracy -50°C to -20°C (-58°F to -4°F) Not specified -19.9°C to -1°C (-3.9°F to 30°F) ± (2% rdg + 6°F/3.3°C) o o -0.9°C to 100°C (30.1°F to 212°F) ± (1.0% rdg + 4°F/2.2°C): 100.1°C to 454°C (212.1°F to 850°F) ± (2.5% rdg + 4°F/2.2°C) 454.1°C to 1000°C (850.1°F to 1832°F) ± (2.5% rdg + 6°F/3.
General Specifications Display 2.2” Color TFT LCD (320 x 240 pixels) Digital Camera 640 x 480 pixels Operating Temperature 32°F to 122°F (0°C to 50°C) Memory Internal flash: 49Mbyte, Micro SD card: Max 8Gbytes Operating Humidity Max. 90% RH (non-condensing) Power Supply 3.7V Lithium ion rechargeable battery Battery life 4 hours (continuous) approx. Battery charge time 2 hours with AC adaptor or USB connection Automatic Power Off Programmable: OFF, 3, 15, and 60 minutes Weight 0.9 lbs.