User`s guide
CO250-EU-EN v1.5 12/13
5
SETUP (alarm limit and temperature scale)
In the normal mode press and hold the
SET
button for more than 1 sec to enter the setup mode.
To exit the setup mode, press the CAL/Esc button when either P1.
0 or P3.0 is displayed.
P1.0 CO
2 ALARM Limit
When entering setup mode, P1.
0 and “AL” are displayed.
Press the M
N/AV button to scroll to P1.1 for setting CO2 alarm threshold. The current CO2 set
value will be blinking.
Press the MODE/▲ button to increase or the DP/WBT/▼ button to decrease the value. Each
press adjusts 100 ppm. The alarm range is from 100 to 9900ppm. When the preferred alarm
value is set, press the M
N/AV button to exit and save the setting or the CAL/Esc button to exit
without saving and return to P1.
0.
P3.0 TEMPERATURE SCALE
Press the MODE/▲ button or the DP/WBT/▼ button in P1.
0 to access P3.0 for setting the
temperature scale.
Press the M
N/AV button to go into P3.1 for setting the temperature units. The currently selected
units (°C or °F) will be blinking in the display. To switch units, press the MODE/▲ button. Press
the M
N/AV button to save the setting or press the CAL/Esc button exit without saving and return
to P3.
0.