Specifications

CG204-EU-EN v2.2 8/13
9
Zerocalibration
1. Placethemeteronanuncoatedsectionofthematerialtobe
measuredoronthereferencesubstrateprovided.Useeitherthe
FerrousorNonFerrousreferenceasrequiredbythemeasurement
application.
a. PlacetheprobeontheuncoatedsubstrateandwatchtheLCD
readings.
b. Whenthereadings
appearstable,liftthemeteroffthesubstrate
c. PressandHoldtheZerobuttonfor2seconds(beep).TheZero
calibrationisnowcomplete.
2. Themeterisnowreadyformeasurements
Calibration
Themeter’saccuracycanbeenhancedbyperformingacalibrationusingthereferencefilmssupplied.
1. Performazerocalibrationasstatedabove
2. Performarangecalibrationusingareferencefilm.
a. Placeoneofthereferencefilmsonthesubstrate.
b. Placethemeteronthecalibrationreferencefilm
and
watchtheLCDreadings.
c. Whenthereadingsappearstable,liftthemeteroffthe
substrate
d. PressMENUandthenscrolldownto
Calibration”.
e. PressSelecttohighlighttheEnablefunctionand
thenpressSelect”.
f. PressEscand
themeterwillenterthecalibration
mode.
g. Adjustthereadingtomatchtheknownvalueofthereferencefilmusingtheand
buttons.
h. TurnthemeterOFFtoexitandsavethecalibrationdata
3. Themeterisnowcalibratedtothebasematerialandthicknessused.
4. PerformStep2asneededusingtheotherreferencefilms.
Notes:
1. CalibrationdataisstoredintomemoryitisnoterasedwhenthemeteristurnedOFF.
2. Frequentlywipethesensorcontactwithacleanlintfreeclothtoremoveanyparticlesonthesensor
tip.
3.
Duringcalibrationtheandbuttonsadjusttheinternalcalibrationfactorswithhighresolution.
Thedisplayresolutionmaybesuchthatitmaytakeupto10pressestoseea1digitchangeonthe
display.
MultipointCalibration
Thismethodrequiresperformingtwoormoresequentialsinglepointcalibrations.Forbestresults,the
expectedcoatingthicknessmeasurementsshouldbewithinthecalibrationpoints.