Specifications
CG204-EU-EN v2.2 8/13
4
ProgrammingMenu
Themetercanbeconfiguredandcalibratedthroughsimplebuttonpressesintheprogrammingmenu.Press
theMENUbuttontoaccessthemenuandrefertothemenu‘tree’below.ThemenuusesUP/DOWN,
SELECT,BACK,&ESCpressesfornavigationandselection.Inthetablebelowthefactorydefaultsettings
are
inboldwithanasterisk.Eachparameterisexplainedindetailinthesubsequentsections.
Toplevel Sublevel1 Sublevel2 Notes
AVG* Averageofaseriesofreadings
MIN Lowestofaseriesofreadings
MAX Highestofaseriesofreadings
NONumberofsampledreadings
STATISTICALVIEW
SDEV Standarddeviationofaseries
Single* Onereadingatatime
Measuremode
Continuous Continuousmeasurements
Direct* Readingsarenotstoredingroups
Workingmode
Group1…4 Storereadingsingroups
Auto* Meterautomaticallyselectsmode
Fe Ferrousmeasurementmode
Probeused
NoFe Non‐ferrousmeasurementmode
µm*
Micrometers
mils Mils=mm*2.54/100
Unitsettings
mm Millimeters
ON* Enablesbacklightoperation
Backlight
OFF Disablesbacklightoperation
AVG* Averageofaseriesofreadings
MAX Lowestofaseriesofreadings
MIN Highestofaseriesofreadings
LCDStatistics
SDEV Standarddeviationofaseries
Enable* Allowsautopowerofftooperate
OPTIONS
AutoPowerOFF
Disable Defeatstheautopowerofffeature
HighLimit HighAlarmalertsuserwhenreached
Limitsettings
LowLimit LowAlarmalertsuserwhenreached
LIMIT
DeleteLimits Clearthealarmlimitvalues
CurrentData Deletecurrentdata
AllData Deleteallstoreddata
DELETE
GroupData
Deletedatastoreddataplusalarmand
calibrationdata
MEASUREMENTVIEW
Viewstoreddatainallgroups
Enable Allowcalibrationaccess
Disable Lockoutthecalibrationmode
DeleteZeroN Clearthezerocal.data(non‐ferrous)
CALIBRATION
DeleteZeroF Clearthezerocal.data(ferrous)
NOTE:DisabletheAutoPowerOFFfeaturebeforelengthyprogrammingtoavoidinconvenientautomatic
powerdownwhileprogramming.