Specifications

CG204-EU-EN v2.2 8/13
4
ProgrammingMenu
Themetercanbeconfiguredandcalibratedthroughsimplebuttonpressesintheprogrammingmenu.Press
theMENUbuttontoaccessthemenuandrefertothemenu‘tree’below.ThemenuusesUP/DOWN,
SELECT,BACK,&ESCpressesfornavigationandselection.Inthetablebelowthefactorydefaultsettings
are
inboldwithanasterisk.Eachparameterisexplainedindetailinthesubsequentsections.
Toplevel Sublevel1 Sublevel2 Notes
AVG* Averageofaseriesofreadings
MIN Lowestofaseriesofreadings
MAX Highestofaseriesofreadings
NONumberofsampledreadings
STATISTICALVIEW
SDEV Standarddeviationofaseries
Single* Onereadingatatime
Measuremode
Continuous Continuousmeasurements
Direct* Readingsarenotstoredingroups
Workingmode
Group1…4 Storereadingsingroups
Auto* Meterautomaticallyselectsmode
Fe Ferrousmeasurementmode
Probeused
NoFe Nonferrousmeasurementmode
µm*
Micrometers
mils Mils=mm*2.54/100
Unitsettings
mm Millimeters
ON* Enablesbacklightoperation
Backlight
OFF Disablesbacklightoperation
AVG* Averageofaseriesofreadings
MAX Lowestofaseriesofreadings
MIN Highestofaseriesofreadings
LCDStatistics
SDEV Standarddeviationofaseries
Enable* Allowsautopowerofftooperate
OPTIONS
AutoPowerOFF
Disable Defeatstheautopowerofffeature
HighLimit HighAlarmalertsuserwhenreached
Limitsettings
LowLimit LowAlarmalertsuserwhenreached
LIMIT
DeleteLimits Clearthealarmlimitvalues
CurrentData Deletecurrentdata
AllData Deleteallstoreddata
DELETE
GroupData
Deletedatastoreddataplusalarmand
calibrationdata
MEASUREMENTVIEW
Viewstoreddatainallgroups
Enable Allowcalibrationaccess
Disable Lockoutthecalibrationmode
DeleteZeroN Clearthezerocal.data(nonferrous)
CALIBRATION
DeleteZeroF Clearthezerocal.data(ferrous)
NOTE:DisabletheAutoPowerOFFfeaturebeforelengthyprogrammingtoavoidinconvenientautomatic
powerdownwhileprogramming.