User`s guide
DO700 V1.0 8/09
19
P4 - Temperature unit of measure selection
• Use the CAL or the M+/RM keys to select the desired unit of measure (
o
C or
o
F).
• Momentarily press MODE to move to the next parameter or press ENTER to return
to the normal measurement mode.
P5 - Display Backlight Setting
• Use the CAL or the M+/RM keys to select 0, 1, 3, or 6 minute default backlighting
time.
• Momentarily press MODE to move to the next parameter or press ENTER to return
to the normal measurement mode.
P6 - Automatic Power OFF Setting
• Use the CAL or the M+/RM keys to select a 0, 10, or 20 minute Auto Power OFF
time.
• Momentarily press MODE to move to the next parameter or press ENTER to return
to the normal measurement mode.
Note: The default setting for DO is 0, that is, the auto power off function is defeated.
P7 - Restore Factory Default Settings
• Use the CAL or the M+/RM keys to select ON (reset the factory default settings) or
OFF (cancel edit).
• Momentarily press MODE to move to the first parameter (P1) or press ENTER to
return to the normal measurement mode.
Replacing the Membrane Cap
Replace the Membrane cap under the following conditions:
• Electrode response time increases.
• Displayed value appears incorrect or abnormal.
• Damaged or cracked membrane.
Replace as follows:
1. Loosen and then remove the membrane cap.
2. Clean the electrode (without membrane cap) with distilled water and shake the
water off of the electrode.
3. Lightly polish the cathode surface (gold) with a piece of clear velvet or tissue.
4. Slowly inject electrolyte into a new membrane cap. After injecting, check the
electrolyte for air bubbles. If air bubbles are present, carefully tap the
membrane cap to remove the air pockets.
5. Place the membrane cap on a clean surface and place the electrode in the cap
vertically. Slowly twist in a clockwise motion until electrode is tightly attached.
Excess electrolyte may be displaced, wipe excess electrolyte with tissue and
distilled water.
6. Check that there are no air bubbles present in the electrolyte (smaller air
bubbles are acceptable). If large air bubbles are present, re-assembly will be
required.
7. Do not touch the sensitive membrane when using the electrode or replacing the
membrane cap. Oils from the skin will affect the quality of the membrane and
decrease the rate of oxygen permeability.