User Manual
MA61_MA63-en-GB_V1.0 3/15 
8
ACCurrentMeasurements
WARNING:Toavoidelectricshock,disconnectthetestleadsfromthemeterbeforemaking
currentmeasurements.
CAUTION:ObserveCATII600V,CATIII300VwithrespecttoEarthGroundfortheJaw.
1.  SettheFunctionswitchtothe600mA,6000mA,or60Arange.
2.  FortheMA63,shortpresstheMODEbuttontoSelectAC,Frequency(orDC)
3.  PresstheTriggertoopentheclampjaw.
4.  Clampontoasingleconductor(fullyenclosingit).Donotallowagapbetween
thetwohalves
ofthejaw.RefertoFig‐3earlierintheOperationsection.
5.  Whenmeasuringvariablefrequencydevices,pressandholdtheVFCbuttontostart/stop
measuringwiththelowpassfilter.
6.  ReadtheACAvalueontheLCD.
DCCurrentMeasurements(MA63only)
WARNING:Toavoidelectricshock,disconnectthetestleadsfromthemeterbeforemaking
currentmeasurements.
CAUTION:ObserveCATII600V,CATIII300VwithrespecttoEarthGroundfortheJaw.
1.  SettheFunctionswitchtothe600mA,
6000mA,or60Arange.
2.  UsetheMODEbuttontoselectDC.
3.  PresstheZERObuttonseveraltimestonull
themeterdisplaywithoutaconductorin
theclampjaw.Zerothemeterwiththe
clampheldinthesameorientationas
the
actualmeasurement(seeFig‐4).
4.  PresstheTriggertoopenthecurrentsense
Jaw.
7.  Fullyencloseasingleconductor.Donot
allowagapbetweenthetwohalvesofthe
jaw.RefertoFig‐3earlierinthisguide.
5.  ReadtheDCAvalueontheLCD.
Fig‐4CorrectandIncorrectZERO
techniques
ZERO
ZERO