User Manual
MA61_MA63-en-GB_V1.0 3/15
8
ACCurrentMeasurements
WARNING:Toavoidelectricshock,disconnectthetestleadsfromthemeterbeforemaking
currentmeasurements.
CAUTION:ObserveCATII600V,CATIII300VwithrespecttoEarthGroundfortheJaw.
1. SettheFunctionswitchtothe600mA,6000mA,or60Arange.
2. FortheMA63,shortpresstheMODEbuttontoSelectAC,Frequency(orDC)
3. PresstheTriggertoopentheclampjaw.
4. Clampontoasingleconductor(fullyenclosingit).Donotallowagapbetween
thetwohalves
ofthejaw.RefertoFig‐3earlierintheOperationsection.
5. Whenmeasuringvariablefrequencydevices,pressandholdtheVFCbuttontostart/stop
measuringwiththelowpassfilter.
6. ReadtheACAvalueontheLCD.
DCCurrentMeasurements(MA63only)
WARNING:Toavoidelectricshock,disconnectthetestleadsfromthemeterbeforemaking
currentmeasurements.
CAUTION:ObserveCATII600V,CATIII300VwithrespecttoEarthGroundfortheJaw.
1. SettheFunctionswitchtothe600mA,
6000mA,or60Arange.
2. UsetheMODEbuttontoselectDC.
3. PresstheZERObuttonseveraltimestonull
themeterdisplaywithoutaconductorin
theclampjaw.Zerothemeterwiththe
clampheldinthesameorientationas
the
actualmeasurement(seeFig‐4).
4. PresstheTriggertoopenthecurrentsense
Jaw.
7. Fullyencloseasingleconductor.Donot
allowagapbetweenthetwohalvesofthe
jaw.RefertoFig‐3earlierinthisguide.
5. ReadtheDCAvalueontheLCD.
Fig‐4CorrectandIncorrectZERO
techniques
ZERO
ZERO