Instruction manual
8 382860 Ver. 1.5 2/01
• MEM (Memory): Up to 10 measurements (0 to 9) can be stored and recalled.
Press the FUNCTION key five times then press the SET/RESET key once to store
the present measurement value in the first available memory location. The memory
location number will flash to the upper left of the left-most sub-LCD. The
memorized reading is shown on the center sub-LCD. Use the UP/DOWN keys to
scroll to the next free memory location. Take another measurement and press the
SET/RESET key to store it. If a memory location number which has been used
before is selected, the previous value is updated (erased) with the new
measurement value. If you accidentally exit the function mode by pressing the
SET/RESET key too many times, the memory is not erased. It can be “recalled”
using the RECALL function which follows…
• RCL(Recall): This function recalls memorized readings to the LCD. Press the
FUNCTION key 6 times until the RCL icon flashes. Press the SET/RESET key
once. The memory location reference number flashes to the upper left of the left-
most sub-LCD. Press the UP or Down keys to select a memory address (a number
from 0-9 where the measurement has been stored). Press the SET/RESET key
again to recall the reading memorized in the selected location. The recalled value
(ref. No. 0) will appear on the left sub-display; the 2
nd
recalled value (ref. No. 1)
appears on the center sub-display and the 3
rd
recalled value (ref. No. 2) appears on
the right side sub-display, and so on for the remaining locations.
• CMP (Comparison): Comparison mode allows readings to be compared to user
programmable High/Low setpoints. Press the FUNCTION key and then the DOWN
key. Press the SET/RESET key once and the word “LOW” should begin flashing
above the left-most sub-LCD. Press the SET/RESET to begin editing the LOW user
setpoint. Use the SET/RESET key to move from digit to digit and use the
UP/DOWN keys to change a digit’s value. When done, press the SET/RESET key
until the LOW setpoint value ceases flashing. The word “HIGH” will begin flashing
over the right-most sub-LCD. Edit the HIGH setpoint in the same manner as the
LOW setpoint. After both reference values have been defined press the
SET/RESET key again. The R-H (Range Hold) icon will begin flashing. Set the
measurement range via the UP/DOWN keys. On the center sub-display “LO”
appears for readings less than the low reference value, “Hi” is displayed for
readings higher than the highest reference value, and “Pass” is displayed for
readings in the region between the low and high reference values. To exit CMP
mode, press the FUNCTION key.
• AUTO-POWER-OFF extends the life of the battery: if neither the rotary function
switch nor any button is operated for 12 minutes the DMM is automatically turned
off. During active “communication” of the multimeter with a PC, AUTO-POWER-
OFF is disabled.