Relative / % Function
The REL/% mode allows for % deviation measurements from a stored reference value.
1. Setup the meter with the parameters required for the test to be performed.
2. Insert the reference component into the test fixture and wait for a stable reading.
3. Press the REL/% button to store the value. The icon will appear in the display.
4. For all subsequent measurements, readings on the lower display will indicate the %
difference between the currently measured component and the stored value.
5. Press and hold the REL/% button for >2 seconds to exit the mode.
Sorting
The sorting mode is used to select components within a % limit of a reference value.
1. Insert the reference component and set the test parameters. “LCR” auto-parameter is not
allowed in the sorting mode.
2. With the desired reading in the display, press the SORT button to establish the reference
value. The main display will indicate “PASS” and the lower display will indicate the value of
the component. The default sorting % is +/- 1%.
Note: If the reference value is above 2000 counts or below 200 counts, sort will not work.
3. Press the SETUP button in sequence to change the Range, Reference Value, and the %
Tolerance.
a. Range: With the “RANGE” icon flashing, press the or button to change the range.
Press the ENTER button to store the setting and proceed to the Value setting.
b. Value Setting: Press the or button to select the flashing digit for adjustment. Press
the or button to adjust the value of the digit. Press the ENTER button to store the
setting and proceed to the Tolerance setting.
c. Tolerance setting: Press the or button to step through the available tolerance
selections:
± 0.25%
± 0.5%
± 1%
± 2%
± 5%
± 10%
± 20%
+80% -20%
Press the ENTER button to store the Tolerance setting.
4. Press the SORT button for > 2 seconds to exit the Sorting mode.