User Manual
IR267‐en‐US_v1.21/17
10
BatteryReplacement
Whenthebatteryiconflashes,orwhenthemeterdoesn’tswitchON,replacethebattery:
Thebatterycompartmentislocatedinthehandgrip.
Thebatterycompartmentlidislocatedjustbelowthetrigger.
Openthebatterydoorbypryingwithathumborfingeratone
oftheindentationsnearthetopofthehandgrip.
Thedoorishingedatthebottomofthehandgrip.
Thebattery
shouldbeinstalledwiththepositive+terminal
(smallercontact)facingthemetertoensurecorrectpolarity.
Safety:Pleasedisposeofbatteriesresponsibly;neverdisposeinafire,batteriesmayexplodeor
leak.Ifthemeterisnottobeusedfor60daysormore,removethebatteryandstoreseparately.
Specifications
TemperatureSpecifications
TemperatureMode Object(SurfaceMode) Ambient(Air) Thermocouple(TypeK)
Range‐58~1112°F(‐50~600°C)‐4~158°F(‐20~70°C)‐58~1832°F(‐50~1000°C)*
Accuracy ±4.1°F(5°C)from‐58~5°F(‐50°C~‐15°C)
±3.9°F(4°C)from5~32°F(‐15°C~0°C)
±3.6°F(2°C)from32~212°F(0°C~100°C)
±3%rdg+1°Cfrom212~1112°F(100°C~600°C)
Accuracyisspecifiedforthefollowingambienttemperaturerange:73~77°F(23~25°C)
Resolution 0.1°F/°C
Emissivity 0.05~1.00adjustable
FieldofView D/S=Approx.12:1ratio(D=distance;S=spotortarget)
Laserpointer Class2laser<1mWpower;Wavelengthis630~670nm
IRSpectral
response
8~14µm(wavelen gth)
*Thesuppliedthermocoupleisratedfor482
o
F(250
o
C)maximumtemperature.Tomeasurehigher
temperatures,pleaseobtainaType‐Kthermocoupleratedforthedesiredtemperature.
GeneralSpecifications
Display Backlitmultifunction LCD display
Displayupdaterate <1secondapprox.
OperatingTemperature 32
o
F~122
o
F(0
o
C~50
o
C)
OperatingHumidity 10~80%RelativeHumiditymax.
StorageTemperature‐14~140°F(‐10
o
C~60
o
C)
PowerSupply One9Vbattery
AutomaticPowerOff Metershutsoffautomaticallyafter10seconds
Weight 4.88oz./138g
Dimensions 6.0x4.7x1.7”(153x120x42mm)