User`s guide
EX470-EU-EN-V5.4-5/12 
14
Function Range Resolution   Accuracy 
5.000Hz 0.001Hz 
50.00Hz 0.01Hz 
(1.5% reading + 5 digits) 
500.0Hz 0.1Hz 
5.000kHz 0.001kHz 
50.00kHz 0.01kHz 
500.0kHz 0.1kHz 
(1.2% reading + 2 digits) 
5.000MHz 0.001MHz 
10.00MHz 0.01MHz 
(1.5% reading + 4 digits) 
Frequency 
Sensitivity: 0.8V rms min. @ 20% to 80% duty cycle and <100kHz; 5Vrms min @ 20% to 80% duty cycle and > 
100kHz
. 
0.1 to 99.9%  0.1% 
(1.2% reading + 2 digits) 
Duty Cycle 
Pulse width: 100µs - 100ms, Frequency: 5Hz to 150kHz 
Note: Accuracy specifications consist of two elements: 
 (% reading) – This is the accuracy of the measurement circuit. 
 (+ digits) – This is the accuracy of the analog to digital converter. 
Diode Test  Test current of 0.3mA maximum, open circuit voltage 1.5V DC typical 
Continuity Check  Audible signal will sound if the resistance is less than 150 (approx.), test 
current <0.7mA 
Temperature Sensor  Requires type K thermocouple 
IR Spectral response  6 to 16µm 
IR Emissivity  0.95 fixed 
IR distance ratio  8:1 
Input Impedance >7.5MΩ (VDC & VAC) 
AC Response  True rms 
ACV Bandwidth  50Hz to 1kHz 
Crest Factor   <3:1 at full scale & <6:1 at half scale 
Display  4000 count backlit liquid crystal 
Overrange indication “OL” is displayed 
Auto Power Off  15 minutes (approximately) 
Polarity  Automatic (no indication for positive); Minus (-) sign for negative 
Measurement Rate  2 times per second, nominal 
Low Battery Indication  “ ” is displayed if battery voltage drops below operating voltage 
Battery  One 9 volt (NEDA 1604) battery 
Fuses  mA, µA ranges; 0.5A/250V fast blow 
A range; 20A/250V ceramic fast blow 
Operating Temperature   5ºC to 40ºC (41ºF to 104ºF) 
Storage Temperature -20
o
C to 60
o
C (-4
o
F to 140
o
F) 
Operating Humidity  Max 80% up to 31ºC (87ºF) decreasing linearly to 50% at 40ºC (104ºF) 
Storage Humidity  <80% 
Operating Altitude  2000 meters.(7000ft) maximum. 
Weight  342g (0.753lb) (includes holster). 
Size  187 x 81 x 50mm (7.36” x 3.2” x 2.0”) (includes holster) 
Safety  For indoor use and in accordance with the requirements for double 
insulation to IEC1010-1 (2001): EN61010-1 (2001) Overvoltage Category 
III 600V and Category II 1000V, Pollution Degree 2. 
Patent Notice  U.S. Patent 7056012 